K22680 Grease Mobility Test Apparatus

Test Method

Determines the resistance of lubricating grease to flow under prescribed conditions. Mobility is measured in grams per second by pumping the sample through a standardized SOD pressure viscometer at controlled temperature and pressure.

Features and Benefits

  • U.S. Steel Method; ASTM Draft Method
  • Test temperatures as low as –30°F (–34.4°C)

K22680 Grease Mobility Test Apparatus

Performs grease mobility tests at low temperatures to predict pumpability characteristics. Determines the suitability of greases for applications in centralized or bulk systems where pumps, valves or pipes are used to distribute or transfer grease. Consists of pressure viscometer, cooling bath and refrigeration system. The stainless steel pressure viscometer is fitted with a modified No.1, 40:1 ratio capillary. After the sample is loaded in the pressure viscometer, the assembly is installed in the cooling bath and allowed to reach the test temperature. Mechanically refrigerated cooling bath can attain test temperatures as low as –30°F (–34.4°C) with stability of ±2°F (±1°C). With the sample at the test temperature, the flow of grease is started under the selected pressure on a nitrogen tank regulator. Flow per second is determined by collecting the grease for a specified period. Includes sample collector turntable.

Specifications

Conforms to: U.S. Steel Method;ASTM Draft Method

Minimum Temperature: –30°F (–34.4°C)

Control Stability: ±2°F (±1°C)

Electrical Requirements

115V 60Hz, Single Phase, 6A
220-240V 50 or 60Hz, Single Phase, 3A

Included Accessories

Grease Cylinder (pressure viscometer) with modified No.1, 40:1 capillary
Sample Collector Turntable

Dimensions lxwxh,in.(cm)

Cooling Chamber: 12x12x30 (30.5×30.5×76)
Refrigeration Unit: 15x12x12 (38x30x30)
Net Weight: 114 lbs (51.7kg)

Shipping Information

Shipping Weight: 188 lbs (85.3kg)
Dimensions: 18.4 Cu. ft.

Ordering Information

Catalog
Number
Product Description Electrical Requirements
K22680
Grease Mobility Appparatus

115V 60Hz
K22685
Grease Mobility Appparatus

220-240V 50Hz
K22686
Grease Mobility Appparatus

220-240V 60Hz

Accessories

CatalogNumber Product Description
K22680-0-22
Grease Cylinder with plunger and fittings
K22680-0-16
Capillary
250-100-001
Thermometer dial type
Range: -100 to +100°F, with 2°F subdivisions